Programming a Totalizer
Output
Programming Site Data
TTLZR
[F3]
5. The
output,
flow. The meter produces a pulse each time the programmed
amount of flow passes through the pipe. Complete the steps below
to program the totalizer output.
[F1] = +TOTL
a. Press
totalize the reverse flow,
the forward mass flow or
reverse mass flow.
b. Enter a value from 50 to 500,000 μsec for the Minimum Pulse
ON
Time and press
Note: A complete pulse consists of equal amounts of
times. Choose a value that is compatible with the frequency
counter to be used.
c. Enter a value for the number of measurement Units/Pulse and
[ENT]
press
.
[F1]
[F4]
6. Press
-
to set up another totalizer/frequency output or press
[EXIT]
to select another slot for setup.
, issues one pulse per selected volume of
to totalize the forward flow,
[F3] = +MASS
(if available) to totalize
[F4] = -MASS
(if available) to totalize
[ENT]
.
April 2008
[F2] = -TOTL
to
ON
OFF
and
1-21