Time Increment Prompt - GE DigitalFlow GC868 Programming Manual

Panametrics gas clamp-on ultrasonic flowmeter, 1-channel
Hide thumbs Also See for DigitalFlow GC868:
Table of Contents

Advertisement

April 2008

TIME INCREMENT Prompt

Procedure Options
3-6
Regardless of which of the various options were selected during the
previous programming steps, all of the paths converge at this point.
16.Use the [←], [→] and
Increment. The available options are: 5sec, 10sec, 30sec, 1min,
3min, 6min, 12min, 30min and 60 min.
The time increment is the frequency at which the Model GC868 takes
and records data measurements. If any reading takes longer than the
programmed time increment, the log is filled in with the next
consecutive reading. For example, assume that a velocity value of 3
ft/sec is recorded at 12:00:00 in a log with a time increment of five
seconds. If the next reading is 8 ft/sec and it takes the meter 12
seconds to read this value, then both of the missed readings (12:00:05
and 12:00:10) will be filled in with the 8 ft/sec value.
You have completed entering data for a standard log. You can now do
one of the following:
Use [↑] and [↓] to review the previous menu prompts.
[F1]-[F4]
Press
to select one of the
To return to the data screen and begin logging data, press the
key.
Although each log is restricted to only three logged parameters, it is
still possible to log more than three parameters. Simply re-enter the
STD
submenu as many times as necessary to set up additional logs.
Select the other desired parameters, and run these logs simultaneously
with the first log.
[F1]-[F4]
keys to select the desired Time
LOG
submenus.
[ENT]
Logging Data