Fuels - TESTO 320 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 320:
Table of Contents

Advertisement

1. Connect the calibration adapter to the flue gas socket.
2. Enable CO measurement parameter: [OK].
3.
[Edit]
4. Attach the connecting line of the test gas bottle to the
calibration adapter.
5. Apply test gas to the sensor.
6. Start recalibration: [Start].
7. Apply the target value once the actual value is stable
(adjustment): [OK].
-or-
Cancel (no adjustment): [esc].
8. Save changes: [Finished].
6.1.4.

Fuels

The fuel can be selected. The fuel-specific coefficients and limits
can be set.
Call up function:
>
[
Activating fuels:
> Select the fuel → [OK].
-
The fuel is activated and the main menu is opened.
Setting coefficients:
1. Select the fuel → [Coeff.].
2. Select the coefficients: [Edit].
Possibly:
> Enter the password:
[OK].
3. Set values → [OK].
4. Save changes: [Finished].
→ Enter the test gas concentration (nominal value).
In order to maintain the measuring accuracy of the
instrument, the correct fuel must be selected or configured.
Correct representation of measuring results is only assured
if the threshold values for the ideal range of the
corresponding measurement task have been set correctly.
The pre-set threshold values are typical values for the
selected system type and the chosen type of fuel.
→ [OK].
]
Fuels
→ Enter password →
[Enter]
6 Using the product
[Next]
41

Advertisement

Table of Contents
loading

Table of Contents