Fuels - TESTO 330 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 330:
Table of Contents

Advertisement

Performing recalibration / adjustment:
WARNING
Dangerous gases
Danger of poisoning!
> Observe safety regulations / accident prevention regulations
when handling test gas.
> Use test gases in well ventilated rooms only.
1. Connect the calibration adapter to the flue gas socket.
2. Select the parameter: [▲],
3.
[Edit]
4. Attach the connecting line of the test gas bottle to the calibration
adapter.
5. Apply test gas to the sensor.
6. Start recalibration: [Start].
7. Accept the nominal value once the actual value is stable
(adjustment): [OK].
-or-
Abort (no adjustment): [esc].
8. Save changes: [Finished].
Pos: 58 /TD/Produkt verwenden/testo 330/testo 330 Brennstoffe @ 6\mod_1279096803274_79.doc @ 65875 @
6.1.4.

Fuels

The fuel can be selected. The fuel-specific coefficients and limits
can be set.
Apart from the pre-configured fuels, 10 more customer specific
fuels can be configured. Fuel parameter, see
www.testo.com/download-center (registration required).
Calling up the function:
>
[
→ Enter the test gas concentration (nominal value).
In order to maintain the measuring accuracy of the
instrument one must choose or configure the correct fuel.
The limits serve the purpose of configuring the ideal zone of
the flue gas matrix and do not have any influence on the
accuracy of the measuring results.
→ [OK].
]
Fuels
→ [OK].
[▼]