The E-MAG Active is a sealed system, which means that the joint is less susceptible to contamination and also pre­
vents clothing from becoming trapped. Not only is this gentle on your clothing but, above all, offers you enhanced
safety.
Two clearly distinguishable switches on a button control unit allow you to electronically unlock the joint (once) or
engage the lock (permanently). In addition, the mechanical (temporary) unlock feature supports a larger area of
application, for example when cycling. These instructions for use describe the functions and operation of the E-
MAG Active as well as how to charge it and care for it.
Please do not hesitate to ask your orthopaedic technician if you have any technical questions.
2 Safety

2.1 Explanation of warning symbols

Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE

2.2 General safety instructions

CAUTION
Mechanical damage to the product
Injuries due to changes in or loss of functionality
► Use caution when working with the product.
► Check the product for proper function and readiness for use.
► In case of changes in or loss of functionality, discontinue use of the product and have it checked by author­
ised, qualified personnel.
CAUTION
Explosion hazard
Do not throw the battery into fire. Never expose it to heat. Avoid direct sunlight.
CAUTION
Risk of injury in case of continued use after malfunctions
Never use the E-MAG Active knee joint if there is a malfunction. Please contact your specialist workshop immedi­
ately.
CAUTION
Risk of injury due to electromagnetic interference
Malfunctions may occur near high-tension power lines, transmitters, transformers or other sources of strong elec­
tromagnetic radiation (such as security systems for goods in department stores). This can cause the patient to
fall.
Avoid proximity to strong sources of magnetic and electric interference (e.g. transformer stations, transmitters).
CAUTION
Risk of falling due to improper operation of the service switches
The small service switches A and T must not be activated or manipulated. They are exclusively for use by qualified
personnel in your specialist workshop.
CAUTION
Risk of falling while using the control unit
To avoid the risk of falling, you may only activate the locking and sitting function while standing and not while
walking.
CAUTION
Risk of falling due to use of grease (or other lubricants)
This could contaminate the joint and jam the lock. The bearing washers and joint locking mechanism should never
be lubricated.
17B203=* E-MAG Active
Safety