Troubleshooting
Troubleshooting Using Diagnostic Test
Table 2-1
7
8

Test equipment required for diagnostic test

Table 2-2
Table 2-2
Required equipment
Required test equipment
24 inch 50  cable

To Execute the Diagnostic Test

To isolate faulty board assembly in analog section, execute the diagnostic test
in accordance with the following procedure. The test procedure needs to be
performed using a mouse and an external keyboard in addition to the front
panel keys.
To perform the diagnostic test properly, the following conditions must be
met:
Environmental temperature: 23 ºC ± 5 ºC
Do not operate front panel keys, keyboard and mouse during the
diagnostic test. Changing the instrument settings while the diagnostic test
is in progress will cause incorrect test results.
Step 1. Connect a mouse and an external keyboard to the E5061B's rear panel
connector.
Step 2. Press
Step 3. Press Load Project to select Load Project function. "Open" dialog box will be
displayed as shown in
Keysight E5061B Network Analyzers
Diagnostic test group menu
Receiver IF Ranging Test
Gain-Phase Absolute Measurement Test
(Option 3L5)
shows the equipment required for performing the diagnostic test.
key.
Figure
Qty
Recommended model
1
8120-8862, 8120-1839
2-13.
33