Ph Value; General Information - wtw Multi 3420 Operating Manual

Digital meter for digital ids sensors
Hide thumbs Also See for Multi 3420:
Table of Contents

Advertisement

Multi 3420
Temperature
measurement
Preparatory activities
ba75864y18
01/2015
4.7

pH value

4.7.1 General information

You can measure the following parameters:
 pH value [ ]
 Sensor voltage [mV]
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This facilitates interference-free measurements also in
the following cases:
 Measurement in grounded test samples
 Measurement in one test sample with several sensors connected to
one Multi 3420
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample. IDS sensors measure the temperature with
a temperature sensor integrated in the IDS sensor.
When operating a sensor without integrated temperature sensor, e.g.
via an IDS pH adapter, you have the following options to determine the
temperature of the measuring solution:
 Temperature measurement via the integrated temperature sensor of
an IDS sensor.
When the measured temperature value is taken from an IDS sensor
the [TP
] status indicator is displayed in the measuring window of
the IDS pH adapter. In the measuring window of the IDS sensor pro-
viding the measured temperature value the [TP
displayed.
 Manual determination and input of the measured temperature value.
Note
The settings for the temperature can be adjusted in the menu for calib-
ration and measurement settings (siehe section 4.7.3).
Perform the following preparatory activities when you want to measure:
1
Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2
If necessary, select the measured parameter with <M>.
3
If necessary, calibrate or check the IDS pH sensor.
Operation
] status indicator is
38

Advertisement

Table of Contents
loading