Wizard Sensor Setup Procedure; Installing Sensors In Reflect Mount - Siemens SITRANS FSS200 Installation Manual

Ultrasonic clamp-on sensors
Table of Contents

Advertisement

Installing/mounting
4.6 Installation instructions
4.6.3

Wizard sensor setup procedure

Referring to the Operation Instuction manual for your transmitter, use the Wizard sensor
setup procedure to set up the clamp-on sensors for your unit.
4.6.4

Installing sensors in Reflect mount

Reflect mount - Sensor installation using mounting frames and spacer bar
1. Use the Wizard setup procedure to program the transmitter for the sensors that were
selected.
2. After receiving the spacing number index from the transmitter, make a note of the number
displayed then prepare the pipe surface area where the sensors will be mounted.
Before continuing refer to the Reflect mount installation diagram example below.
Figure 4-6
30
Sensor clamping screw
Customer pipe
7ME39600*** Mounting Frame
Spring Clip (not present on some mounting
frame models)
Spacer bar
Reflect mount
Spacer bar pin and reference hole
Mounting strap
Larger pipes may need additional mount-
ing straps
Orientation for Single path sensor at the 9
o'clock position
Orientation for Dual path sensor at the 10
and 2 o'clock positions
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC