Additional Installation Instructions
B.3 Magnetic mounting
B.3.5
Installing the sensors
Sensor installation
1. Take either sensor and apply a continuous lengthwise 3 mm (1/8-inch) bead of coupling
compound across the center of the sensor emitting surface
①
②
③
Figure B-21
2. Slide sensor into the angled end of the Reference mounting frame, with the sensor F-
connector facing out. Keep sensor from making contact with the pipe until it butts up
against the mounting frame stop. Push sensor down to mate with pipe.
Figure B-22
110
④
F-connector
Front face
⑤
Angled edge
Emitting surface
⑥
Sensor
Coupling compound
Sensor
Sensor mounting top locations
①
Stop location for D and E-size FSS200 sensors
②
Stop location for C-size FSS200 sensor
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC