Additional Installation Instructions
B.4 Hi Precision mount
1. Thread unterminated sensor cable through enclosure housing grommet and terminate
cable end with F-connector and cable gland assembly. For sensor cable selection refer to
Technical data (Page 81). For sensor assembly instructions refer to Sensor wiring
(Page 48).
Note
If required, replace supplied grommet with conduit fitting on cable gland.
①
Sensor cable and grommet
②
Cable gland
③
Sensor
Figure B-34
Hi Precision mount Single enclosure sensor installation
2. Fill connector end with Super Lube prior to connecting.
3. Apply Super Lube to the internal threads at the large end of the thread connector.
4. Connect cable to sensor and then thread cable gland on to sensor.
5. Take either sensor and apply a continuous lengthwise 3 mm (1/8-inch) bead of coupling
compound across the center of the sensor emitting surface.
Note
Within the side walls of the enclosure housing there is a short vertical slot and a long
horizontal slot. The short vertical slot is the reference sensor position (REF).
6. Place the spacer bar on the enclosure side wall and insert the screw pin through the REF
hole of the spacer bar.
122
④
Notch
⑤
Clamp plate, screw and nut
⑥
Spacer bar and screw pins
Installation Manual, 08/2017, A5E36255466-AC
FSS200 clamp-on sensors