Required Tools - Siemens SITRANS FSS200 Installation Manual

Ultrasonic clamp-on sensors
Table of Contents

Advertisement

Figure 1-1
1.5

Required Tools

It is helpful to have common electric and hand tools available for installation of the sensors.
● For easy programming, a PC would also be helpful and can be used later for diagnostics
and documentation.
● Sensor cables are normally preassembled at the transmitter end.
● Hot-air gun if shrink sleeves are needed.
● A plastic hammer and bubble level should be used for sensor mounting and Vernier
adjustment of sensor frames.
● For preparing irregular pipe surfaces that have old paint, corrosion, etc. an electric hand
sander may be used.
● An optional circumference tape measure for pipe and sensor installation.
Note
Important
Never use an electric hand grinder or angle grinder when preparing pipe surfaces.
Incorrect bevel and chamfer angles at the pipe surface can disturb clear signal transmission.
● Use a thickness gauge to measure pipe wall thickness. If not available use pipe class
tables or call technical support.
● Assorted sizes of screwdrivers, nut drivers and wrenches.
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC
Mounting frames
Introduction
1.5 Required Tools
9