Connecting
5.5 Connecting sensors to transmitter
Connecting the flow sensor cables
For the system to function properly, the sensor cables must be connected correctly. Each
sensor cable should be connected to its own port assignment.
The sensor cables are delivered with a premounted F-connector at transmitter end.
1. Remove blind plug from wall mount housing.
2. Slide the cable gland back onto the cable and allow access for ther F-connector tool.
3. Place F-connector tool around cable and slide up to engage F-connector nut.
4. Push cable through gland opening. Ensure that center lead is aligned with connector port
in transmitter.
5. Tighten F-connector using tool until you meet mechanical stop.
6. Remove F-connector tool.
7. Mount and tighten cable gland.
Repeat these steps for each sensor cable.
62
1-1A
2-2B
3-2B
4-2A
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC