Connecting
5.4 Sensor wiring
Assemble the cable and F-connector as follows:
4. Push and twist the F-connector
5. Ensure the dielectric
conductor
5.4.2.2

Connecting coaxial cable

1. Take the prepared sensor cable end and fill the F-connector with Super Lube coupling
compound, then screw it onto the sensor
2. Coat the NPT Adapter
3. Screw the cable gland body hex into the NPT Adapter. Tighten the Gland body hex
until it butts up against the NPT Adapter end (approx. 8Nm).
4. Using an adjustable wrench, screw on the gland cap nut
optimum sealing. The gasket
50
is flush to 1mm (.04 in) beyond barrel end and that the center
is 1.5mm (0.06 in) beyond the F-connector body
with Super Lube coupling compound, screw it onto the sensor
and hand tighten.
on the cable until secure.
and hand tighten.
must protrude along the cable as shown.
Installation Manual, 08/2017, A5E36255466-AC
.
and tighten to obtain
FSS200 clamp-on sensors