Siemens SITRANS FSS200 Installation Manual page 101

Ultrasonic clamp-on sensors
Table of Contents

Advertisement

Positioning track assemblies
1. Wrap a length of the Mylar spacing guide around the pipe and against the end of the track
assemblies. Ensure that the spacer guide edges on both sides align. Arrange so that one
end overlaps the other by at least 8 cm (3 inches). Trim to fit if necessary, but in order to
keep the end square, be sure not to trim at the overlapping end.
Figure B-8
2. Remove the spacer guide. Measure or fold spacer guide to find its halfway distance. Mark
a center line and then tape spacer guide to pipe.
Figure B-9
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC
Trim material from inner edge if necessary
Mylar Spacing Guide
8 cm (3-inches) Overlapping Edge
Wrapping the Mylar spacing guide around the pipe (end view)
Overlap Edge
Mark (or fold) exactly at half-way point
Finding the halfway distance
Additional Installation Instructions
B.2 Mounting tracks for Size A and B sensors
Mark on Spacing Guide
Circumference
99