Cat, Na, And Vvm Modes; Sa And Cpm Mode - Keysight Technologies FieldFox N9913A Service Manual

Handheld analyzers
Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

Theory of Operation
Operation in Modes
Modes
CAT/TDR
ERTA Source
Source
Enable
OFF ON
Option 209
& 233
For a given mode, the analyzer "remembers" the most recent state of the
current mode when you change to another mode.
For most screens under System 7 key, the current measurement mode
continues to operate while the screen is displayed. For instance, while the
battery information screen is displayed, the measurement mode remains
active.
There are two exceptions to this:
— the "Options (Licensing)" screen
— "Error Log" screen
When either of these screens is displayed, the firmware is in a special monitor
mode and measurements are suspended.

CAT, NA, and VVM Modes

CAT, NA, and VVM modes have some similar capabilities but are still separate
operating modes.
Measurement similarities:
— Return Loss = S11 magnitude
— Insertion Loss = S21 magnitude
Unlike S11 and S21, the values displayed below zero for Return Loss and
Insertion Loss are not negative.
There is user control of output power level, which is high, low or manual setting
on the RF Out Atten, 35 dB in or out.

SA and CPM Mode

SA Mode measures signals at the SA RF IN connector.
SA Mode measurements require NO calibration.
The signal path for SA mode is the same as the input path for two port CAT/NA
modes.
80
NA
SA
Source Mode
Normalize 
(CW)
(OFF ON)
Option 209
Option 209
& 233
& 233
VVM
Power Meter
(USB Sensor)
Source Power
Source CW
Freq
Option 209
Option 209
& 233
& 233
Keysight FieldFox Handheld Analyzers Service Guide
Power Meter
(Buil t In)