Keysight Technologies FieldFox N9913A Service Manual page 125

Handheld analyzers
Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
4. Carefully remove the serial number label from the inside bottom of the old
5. Refer to
6. Remove the dowel pins (5) from the old front case and install them in the
7. Make sure that the gaskets (6) are all in place in the grooves, that there
8. Refer to
9. Refer to
10.Reinstall the rear case onto the front case. Refer to
Keysight FieldFox Handheld Analyzers Service Guide
front case and apply the label to the new front case.
Figure
7-13. Choose the appropriate model number nameplate
overlay for your analyzer, remove the protective backing to expose the
adhesive, and adhere it in the location shown.
new case. The other two dowel pins should already be installed in the
case.
are no gaps, and that they are not damaged in any way.
"Removing and Replacing the Front Panel Group Assemblies" on
page 130
to reinstall the front panel group assemblies removed in
"Removing and Replacing the Measurement Group Assemblies"
on page 136
to reinstall the measurement group assemblies removed in
step
2.
and Rear Cases" on page
126.
step
3.
"Combining the Front
125

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents