Keysight Technologies FieldFox N9913A Service Manual page 64

Handheld analyzers
Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

Theory of Operation
Measurement Group
SA RF Section
An external RF input signal can be analyzed when connected to the Port 2 SA
RF Input connector. Before the input signal is mixed, the signal will go through
the input attenuator and optional internal preamplifier.
The input attenuator default setting from the factory is 10 dB. The input
attenuator is used to control the amount of power at the input of the 1st
converter. Changing the input attenuator value can improve the dynamic range
of the instrument and reduce internally generated distortion products.
When the optional internal preamplifier is turned on, the displayed average
noise level of the instrument is improved. This occurs because the noise figure
of the overall system is improved. Improving the displayed average noise level
of the system enables the user to analyze low level input signals.
The 1st, 2nd and 3rd converters are used to downconvert the input signal to a
final I.F. of 33.75 MHz. That I.F. is fed to the ADC on the A5, System board. This
final I.F. on the output of the ADC can be measured at the IF Out SMB (m)
connector on the side of the instrument. Refer to
Output" on page
The frequency bands for the SA circuitry on the RF board are listed below:
Table 5-3
64
49.
Frequency Bands - SA RF Section
N991xA, N992xA, N993xA
Band
Instrument
Start
1
0.030
2
15.000
3
500.000
4
1366.250
5
2600.000
6
3900.000
7
7500.000
8
9500.000
9
12300.000
10
15700.000
11
16300.000
12
19500.000
13
23000.000
14
23734.000
15
25500.000
"Verifying the 33.75 MHz I.F.
Stop
15.000
500.000
1366.250
2600.000
3900.000
7500.000
9500.000
12300.000
15700.000
16300.000
19500.000
23000.000
23734.000
25500.000
26500.000
Keysight FieldFox Handheld Analyzers Service Guide