Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
Replacing the Strap Handle
1. Separate the front and rear cases to the positions shown in
2. Note the orientation of the strap handle, then lift it off of the dowel pins at
3. Slip the new strap handle over the dowel pins.
4. Reinstall the rear case onto the front case. See
Replacing the I/O Side Panel, Doors, and Gasket
1. Separate the front and rear cases. See
2. Refer to
120
Separating the Front and Rear
each end.
Rear
Cases.
Cases" on page
115.
Figure
7-10. Disconnect the A5 System board to A3 FPIB cable,
W14, at the A3 FPIB connector. If replacing this cable, disconnect both
ends.
Cases.
Combining the Front and
"Separating the Front and Rear
Keysight FieldFox Handheld Analyzers Service Guide