Keysight Technologies FieldFox N9913A Service Manual page 31

Handheld analyzers
Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

Tests and Adjustments
Before You Begin
Table 3-1
Connector Care Quick Reference Guide
Do
— Keep connectors clean
— Extend sleeve or connector nut
— Use plastic end-caps during storage
Do
— Inspect all connectors carefully
— Look for metal particles, scratches, and
dents
Do
— Try compressed air first
— Use isopropyl alcohol
— Clean connector threads
Do
— Clean and zero the gage before use
— Use the correct gage type
— Use correct end of calibration block
— Gage all connectors before first use
Do
— Align connectors carefully
— Make preliminary connection contact
lightly
— Turn only the connector nut
— Use a torque wrench for final
connection
1. Cleaning connectors with alcohol shall only be done with the instrument's power cord removed, and
in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate
prior to energizing the instrument.
Keysight FieldFox Handheld Analyzers Service Guide
Handling and Storage
Do Not
Visual Inspection
Do Not
Connector Cleaning
Do Not
1
Gaging Connectors
Do Not
Making Connections
Do Not
— Touch mating-plane surfaces
— Set connectors contact-end down
— Store connectors or adapters loose
— Use a damaged connector - ever
— Use any abrasives
— Get liquid into plastic support beads
— Use an out-of-specification
connector
— Apply bending force to connection
— Over tighten preliminary connection
— Twist or screw any connection
— Tighten past torque wrench "break"
point
31