Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
5. Refer to
6. Install the three doors in the new I/O side panel:
Figure 7-12
Replacing the I/O Side Panel Doors and Gasket
7. Before installing the new I/O side panel, or when reinstalling the existing
8. To install the new I/O side panel, place it in position on the A5 System
Keysight FieldFox Handheld Analyzers Service Guide
Figure
7-12. Remove the three doors from the old I/O side panel:
a. Open each door to its stopping point, then GENTLY force it beyond
its stopping point to separate it from the panel.
a. Place each door in its respective opening with the hinges aligned
with the hinge slots in the side panel.
b. Press the hinges into their slots until they snap into place.
I/O side panel, install a new gasket in the groove around the perimeter of
the panel.
If an old gasket must be removed, use tweezers being careful not to
damage the groove.
Start in the groove near the flat end of the side panel. Be careful not to
stretch or damage the gasket in any way. Also, make sure there is no slack
in the gasket. A new gasket is provided in the
(see
page
103). It is the smaller of the two gaskets provided in the kit.
board and reinstall the two attachment screws. See
Repair and Re-assembly Kit
Figure
7-11.
123