Chapter 2. Instrument Setup
2.5.1 1-Point Calibration (PROBE & CAL - CALIBRATION)
To calibrate the DMS Go, follow the instructions in this section.
In the MAIN menu, scroll to the PROBE & CAL submenu using the joystick (
1.
2.
Use the joystick (
Start.
3.
For 1-point calibration, verify the thickness of the calibration block/piece with the value displayed on the DMS Go when in calibration
mode (if applicable - after probe zeroing has been completed). Press F2 (ENTER) to manually change the calibration value. If the
calibration block/piece value is correct, couple the probe to the thickness calibration block. The DMS Go will automatically perform
the calibration and adjust the material velocity to match that of the calibration block/piece.
62
Your Complete Source for
Testing Equipment. Since 1969!
) to select the function titled CALIBRATION. Then, press the function toggle or move the joystick (
www.BergEng.com
Berg Engineering & Sales Company, Inc.
).
) to
DMS Go Operator's Manual