Hydran* M2-X Instruction Manual
8 Troubleshooting
WARNING
The following useful references help to understand the instructions given in this
Chapter:
• Alarms operation (see Chapter 6).
• Display of Hydran M2-X's alarm messages (see Section 3.1.2.1).
8.1 Alarm Messages that can be Displayed due to System
Fault Conditions
Appendix I presents the list of all alarm messages that can be displayed by the
Hydran M2-X. This Section contains, in alphabetical order, all alarm messages that
can be displayed due to system fault conditions.
8.1.1
Battery L-Low Alarm
• Alarm type: Due to system fault conditions.
• Non-abbreviated description: Low-Low battery voltage.
• Alarm message on the Perception software: Battery Low-Low.
• Priority: Average.
• Parameter(s) involved:
• Batt. LL; SysOK Relay;FaultTrig. submenu; see Section 3.2.3.
• Batt.Alr LL (alarm set point); Service;VoltPile submenu; see Section 3.3.2.3.
• BattAlr Delay; Service;VoltPile submenu; see Section 3.3.2.4.
• Alarm cause(s): Battery voltage below the Alarm Lo-Lo SP alarm set point.
Replace the battery within the next three months. For details, see Appendix G.
8.1.2
Battery Low Alarm
• Alarm type: Due to system fault conditions.
• Non-abbreviated description: Low battery voltage.
• Alarm message on thePerception software: Battery Low.
• Priority: Very low.
• Parameter(s) involved:
MA-029
Rev. 1.0, 30-Jul-18
Read all warnings and recommendations in Chapter 2 before
attempting one of the troubleshooting procedures listed in
this Chapter.
Page 125