Calibration Signal For Load Cells - Instron 4400 Series Operator's Manual

Universal testing system
Table of Contents

Advertisement

Strip Chart Recorder
Table 6-3.
LOAD CELL
MAXIMUM
CAPACITY
50 N
500 N
5 kN
10 kN
50 kN
150 kN
300 kN
600 kN
Note
6-10
(1) To close the load cell relay, press the LOAD CAL
key on the Series 4400 front panel, then press 0 and
ENTER. The LOAD CAL and LOAD BAL LEDs
are both lit when this relay is closed.

Calibration Signal for Load Cells

*CALIBRATION SIGNAL VALUE - SI EQUIVALENT
NEWTONS
POUNDS
25
250
2 500
5 000
1 124
25 000
5 620
75 000
16 861
150 000
33 721
300 000
67 443
*To convert newtons:
1. Multiply by 0.224 808 9
to obtain pounds.
(2) IMPORTANT – After checking pen deflection,
open the calibration relay by repeating the key se-
quence to close the relay: press LOAD CAL, 0,
ENTER on the keypad. The LOAD CAL and
LOAD BAL LEDs go out. A test cannot be
started when the calibration relay is closed.
The voltage generated when the load cell cali-
bration relay is closed is the same as would be
produced by an applied load equal to 50% of
the maximum capacity of the cell when using
M10-94400-1
KILOGRAMS
5.62
2.55
56.2
25.49
562
254.9
509.8
2 549
7 648
15 296
30 592