Time Base Operation - Instron 4400 Series Operator's Manual

Universal testing system
Table of Contents

Advertisement

X-Y Recorder
(g) If a non-self identified transducer is installed, hang a
(h) Adjust the CAL variable control on the recorder to

Time Base Operation

You can use the X-Y Recorder in a time-based mode, if de-
sired. The Y-Axis signal is plotted against a recorder-gener-
ated time base which can be correlated with crosshead
speed. The time scale can be switch-selected for a range of
0.1 seconds per centimeter to 50 seconds per centimeter or
0.25 seconds per inch up to 125 seconds per inch.
Since the Model 5500 uses chart speeds in units of
inches/minute or centimeters/minute, and the Recorder
Time Scale is in units of seconds/inch or seconds/centime-
ter, Table 6-5 gives speed conversions between the two.
Table 6-5.
Sec/In
0.25 0.5
In/Min
240
Sec/Cm 0.1
Cm/Min 600
To set up the Time Base on the Recorder, do the following:
(a) Press the X-Y/Y-T MODE switch down to Y-T.
6-22
precision weight, within the range of the load cell,
from the cell, or displace the extensometer with the
calibrator, and observe the pen deflection.
set the pen to exactly the value of the precision
weight or extensometer deflection.
Chart Speed Conversions
Chart Speeds
1.25 2.5
5
120
48
24
12
0.2
0.5
1
2
300
120
60
30
M10-94400-1
12.5 25
50
125
4.8
2.4
1.2
0.48
5
10
20
50
12
6
3
1.2