When "HELP" is on the display, press [ENTER] to
change the display to "E". This indicates that the Error
type command is in effect.
The "E" type commands are numbered 0 through 9, ex-
cept "4" and "7", which are not used. You must enter
one of these numbers on the keypad, followed by [EN-
TER] to execute an "E" command. E0 lists error mes-
sages in all functional sections, E1 to E8 lists error
messages for sections 1 through 8 (except 4 and 7), re-
spectively, and E9 prints the list of error messages.
Table 8-1 lists the functional sections in the Series 4400
control console.
Table 8-1.

Sections of the Control Console

Section No.
Functional Name
1
IEEE-488 Interface
2
Central Processing Unit (CPU)
3
Crosshead Control
4
5
Load Sensor Conditioner
6
Strain Sensor Conditioner
7
8
Recorder Interface
The following examples show the use of "E" commands
after entering the Diagnostics Monitor as described pre-
viously.
EXAMPLE NO. 1:
To display errors for each Section sequentially, 1
through 8:
(Not Used)
(Not Used)
Error Messages
Basic or Option
Option
Basic
Basic
- -
Basic
Option
Option
8-5