Electrical Calibration Of Self Identifying Strain Gauge Extensometers; Manual Calibration Of Strain Gauge Extensometers - Instron 4400 Series Operator's Manual

Universal testing system
Table of Contents

Advertisement

Strain Calibration
Electrical Calibration of Self Identifying Strain
Gauge Extensometers
Note
To electrically calibrate a 2630 Series self-identifying
strain gauge extensometer:
(a) Clamp the extensometer onto the specimen at gauge
(b) Press the STRAIN CAL key. The Main Panel Dis-
(c) Press ENTER. After about 6 seconds, the STRAIN

Manual Calibration of Strain Gauge Extensometers

To manually calibrate a 2630 Series self-identifying, or
non-self identifying, strain gauge extensometer:
(a) Mount spindles on the calibrator that fit the exten-
(b) Adjust the calibrator for minimum displacement
5-22
When electrically calibrating a 2630 Series
self-identifying strain gauge extensometer, the
balance operation is performed automatically.
length, as described in the manual supplied with the
extensometer. Allow up to 15 minutes for the exten-
someter to stabilize after applying excitation.
play shows the maximum strain (%) or range (mils or
mm) of the extensometer, and the STRAIN CAL
LED lights.
CAL LED goes out, the display goes blank and cali-
bration is completed.
someter clamps, and set the calibrator in a vertical po-
sition for greatest accuracy.
(gauge length). Minimize backlash in the mechanism
by going through the zero point and then return.