Installation

Power Requirements

The X-Y Recorder is set at the factory to accept a main
power input of 120 ±10 Va.c., single phase. A 3-wire
power cable for this voltage is included. If the power
source to be used is not 120 ±10 Va.c., the power input
switch on the underside of the recorder can be set for
other voltages in the range of 100 to 240 Va.c. (refer to
Table 6-2).
To adapt the X-Y Recorder for a different line voltage,
set the input power switch to the proper input voltage for
your location. Be sure the line fuse is the proper value
for the line voltage in use (see the Recorder manual).
If the X-Y Recorder is to be adapted for 200-240 Va.c.
operation, you must alter the power cable by adding a
male plug that is specified for that power source outlet.
Observe the following EEC wire color code for the ca-
ble:
Brown - high (live)
Light Blue - low (neutral)
Green & Yellow - earth/ground

Signal Input Connection

To use the X-Y Recorder with a Series 4400 system,
plug the cable (A570-26) provided with the unit into the
connector on the underside of the recorder and into the
ANALOG OUT connector on the rear panel of the Se-
ries 4400 console.
X-Y Recorder
6-15