Display Section - Instron 4400 Series Operator's Manual

Universal testing system
Table of Contents

Advertisement

Function Of Controls

Display Section

Table 4-2.
CONTROL or
INDICATOR
LOAD
DISPLAY
EXTENSION
DISPLAY
STRAIN
DISPLAY
TRACK
4-12
Figure 4-2.
Display Section
Display Section Functions
FUNCTION
A 4-digit display which indicates + and - values of load
between .0001 and 9999. The display shows "——" when
the load cell is uncalibrated; is blank without data; and
shows "EEEE" when overranged.
A 4-digit display which indicates + and - extension values
between .0001 and 9999 from gauge length. The display
shows "——" if the load frame has not been identified,
and is blank if no data is available.
A 4-digit display which indicates strain values between
.0001 and 9999. The display shows "——" if the strain
channel is not calibrated, "EEEE" if it is overranged, and
is blank if strain is not installed or no data is available.
Sets the displays to show instantaneous values of load,
extension and strain. The displays are updated every 300
msec during a test. The TRACK LED is lit when this key
is active. (Peak and break values are recorded even
though tracking is active.)