Instron 4400 Series Operator's Manual page 158

Universal testing system
Table of Contents

Advertisement

X-Y Recorder
The X-Axis and Y-Axis Modules are essentially identi-
cal and both are calibrated in the same way. Use the fol-
lowing procedure for both types of Modules:
(a) Set the polarity switch (+/–) to your required polarity
X-Axis:
Y-Axis:
(b) Set the pen to the zero position by pressing the O/M
(c) After the pen has been zeroed, press the O/M switch
(d) Set the MAGNIFICATION FACTOR switch to 1.
(e) Set the CAL/VAR switch to VAR.
6-20
as follows:
+ (button UP) = increasing signal toward top of chart
– (button DOWN) = increasing signal toward bottom
of chart
+ (button UP) = increasing signal toward right of
chart
– (button DOWN) = increasing signal toward left of
chart
button so that is pops up to O (Up position). This dis-
ables the input and sets the pen to the edge of the chart.
If the pen is not exactly on the edge of the chart (press
the Pen Lift switch to lower the pen to see where it
touches the paper), use the thumbwheel on the Axis
Module to adjust the pen position. Note that the Y-Axis
thumbwheel will move the pen up and down, while the
X-Axis thumbwheel moves it left and right.
to M. If you fail to do this, you will not be able to
read any input signals.
M10-94400-1