System Configuration
5.
Repeat step
export.
6.
When you are done, press F6 (Return).

Import Setup

The Import Setup utility allows you to import up to five system setup files from another
device that were exported to an SD card. This feature is useful to sites with multiple
systems that need to have the same or similar setups.
1.
Insert the SD card with the saved setup file.
2.
From the Main Menu, press F4 (System Configuration) > F6 (More) > F6 (More) >
F2 (Import).
The Select Setup for Import window opens. All saved settings on the device
are listed in the left column. All saved settings on the SD card are listed in the
right column.
3.
In the right pane, select the setup file you want to import.
4.
Press F1 (Import).
The selected file is copied to the device and is displayed in the left column.
5.
Repeat step
import.
6.
When you are done, press F6 (Return).

Exporting Audit Trail

The Export Audit function copies the system audit trail in XML format to an SD card.
The Audit Trail tracks the creation, transmission, and deletion of records, changes to
the system setup, and tracks the ID of the users who made each change.
Audit trail log files are saved to the audittrail directory on the SD card. Their filenames
are in the format audittrail_x.log, where x is a number. When a log file is saved
to the SD card, the system determines whether the card already contains an audit trail
log file and names the new file accordingly. For example, if the card does not contain a
log file, the new file is named audittrail_0.log; subsequent files are increments
of 1: audittrail_1.log, audittrail_2.log, audittrail_3.log, and
so on.
After the log file is exported to the SD card, it is cleared from the MAC system.
GE Healthcare recommends exporting the audit trail weekly to long term storage to
meet archive requirements. If the audit trail is not exported regularly, it consumes
storage space and reduces the number of ECGs that you can store on the system.
To export an audit trail, the following conditions must be met:
• High Security Mode must be enabled.
See
• Audit Trail must be enabled.
See
• The user must have the Edit Setup and Delete Records permissions set.
See
120
3
through step
3
through step
"Basic Setup" on page
"Basic Setup" on page
"User Setup" on page
113.
4
for each saved configuration file you want to
4
for each saved configuration file you want to
87.
87.