System Errors; Mac™ 800 - GE MAC 800 Operator's Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 800:
Table of Contents

Advertisement

Refer to
• Verify connectivity by doing the following:
• If a wired LAN is used, verify that the network cables are connected.
• If a wireless LAN is used, verify that the WIFI module is inserted in the USB port
on the device, and the wireless networking setting is configured properly.
Refer to
• Verify the IP, netmask, gateway, and DNS server addresses are all correct.
Refer to
• Ping the device from the file server to verify that the two devices can
communicate.
• Verify the FTP server name or IP address, and export path are correct.
• Verify the logon information is correct.
Check the user name and password. Refer to
for information on the logon information.
• Verify FTP server permissions.
Ensure that the account used to log on to the FTP has write permissions to the
FTP path.

System Errors

The following table identifies some potential errors that may occur while you are
operating the system, the possible causes, and a recommended course of action
to resolve the error.
If performing the recommended actions does not resolve the problem, contact
authorized service personnel.
Battery Error message is
displayed.
The battery LED flashes
intermittently when
operating from battery
power.
The following message is
displayed: The writer door is
open.
The system powers down
while operating from battery
power.
The following message
is displayed: Lead
disconnected.
2060026-001C
"Communication Setup" on page 100
"Communication Setup" on page 100
"Communication Setup" on page 100
Problem
The battery is installed
incorrectly, or the battery is
not functioning correctly.
Battery charge is low.
The writer door is closed
incorrectly.
Battery is fully discharged
One or more electrodes is
disconnected.
MAC™ 800
for information on activating options.
for instructions on settings.
for instructions on settings.
"Communication Setup" on page 100
Cause
Verify the battery contacts
are clean.
Notify service to check and
replace the battery.
Connect the system to an
AC wall outlet to charge the
battery.
Close the door correctly.
Connect the system to an
AC wall outlet to charge the
battery or power the device.
Reconnect the electrodes.
Troubleshooting