YOKOGAWA PZ4000 User Manual page 49

Power analyzer communication interface
Hide thumbs Also See for PZ4000:
Table of Contents

Advertisement

4.6 CURSor Group
:CURSor:HORizontal:Y<x>?
Function
H Queries the Y-axis value (physical value) of
the H cursor.
Syntax
:CURSor:HORizontal:Y<x>?
Example
:CURSOR:HORIZONTAL:Y1?→50.000E+00
Description
• "
(Not A Number)" will be returned, if the
NAN
"
:CURSor:TYPE
HAVertical.
• "
NAN
(Not A Number)" will also be returned, if
the "
:DISPlay:FORMat
include the waveform display.
:CURSor:MARKer?
Function
Queries all settings related to the marker.
Syntax
:CURSor:MARKer?
Example
:CURSOR:MARKER?→:CURSOR:MARKER:
TRACE1 1;TRACE2 1;POSITION1 20.000E-03;
POSITION2 80.000E-03
:CURSor:MARKer:DX?
Function
Queries the X-axis value (physical value)
between the markers.
Syntax
:CURSor:MARKer:DX?
Example
:CURSOR:MARKER:DX?→60.000E-03
:CURSor:MARKer:DY?
Function
Queries the Y-axis value (physical value)
between the markers.
Syntax
:CURSor:MARKer:DY?
Example
:CURSOR:MARKER:DY?→157.26E+00
Description
• "
(Not A Number)" will be returned, if the
NAN
"
:CURSor:TYPE
• "
(Not A Number)" will also be returned, if
NAN
the "
:DISPlay:FORMat
include the waveform display.
4-20
" is not set to Horizontal or
" setting does not
" is not set to
.
MARKer
" setting does not
:CURSor:MARKer:FFT<x>
Function
Sets the X-axis value of the marker position for
the FFT result or queries the current setting.
Syntax
:CURSor:MARKer:FFT<x> {<frequency>|
<NRf>}
:CURSor:MARKer:FFT<x>?
<x> = 1 to 2
<frequency> = 0 to 2.5 MHz (normal
<NRf> = 0 to 5000 (when Time Base =
External or during harmonic
measurement)
Example
:CURSOR:MARKER:FFT1 200kHz
:CURSOR:MARKER:FFT1?→:CURSOR:MARKER:
FFT1 200.0E+03
Description
• This command is valid when
"
:CURSor:MARKer:TRACe<x>
MATH<x> and the equation of MATH<x> is
set to FFT.
• The range and resolution of <frequency> is
determined from the sampling rate and the
number of FFT points.
• <NRf> is set in terms of harmonic order. The
range depends on the number of FFT points
as follows. For the procedure to set the
number of FFT points, see the
"
:MATH<x>:FFT:POINt
For 1000 points : 0 to 500
For 2000 points : 0 to 1000
For 10000 points : 0 to 5000
:CURSor:MARKer:JUMP
Function
Jumps to a waveform of the marker.
Syntax
:CURSor:MARKer:JUMP {M1_MA|M1_Z1|M1_Z2|
M2_MA|M2_Z1|M2_Z2}
Example
:CURSOR:MARKER:JUMP M1_Z1
Description
The parameters "
markers 1 and 2, respectively. "
"
" represent the main waveform, zoomed
Z2
waveforms 1 and 2, respectively.
:CURSor:MARKer:PERDt?
Function
Queries the 1/∆ value of the horizontal axis
between the markers.
Syntax
:CURSor:MARKer:PERDt?
Example
:CURSOR:MARKER:PERDT?→16.667E+00
measurement, when Time
Base = Internal)
" is set to
" command.
M1
" and "
M2
" represent
MA,
" "
Z1,
" and
IM 253710-11E

Advertisement

Table of Contents
loading