YOKOGAWA PZ4000 User Manual page 93

Power analyzer communication interface
Hide thumbs Also See for PZ4000:
Table of Contents

Advertisement

4.15 NUMeric Group
:NUMeric:HARMonics:PRESet
Function
Sets the numerical data output items to a preset
pattern during harmonic measurement.
Syntax
:NUMeric:HARMonics:PRESet {<NRf>}
<NRf> = 1 to 4
Example
:NUMERIC:HARMONICS:PRESET 1
Description
For information related to the output items that
are set to preset values, see "
numerical data output items that are
" on page 4-67.
preset
:NUMeric:HARMonics:VALue?
Function
Queries the numerical data during harmonic
measurement.
Syntax
:NUMeric:HARMonics:VALue? {|<NRf>}
<NRf> = 1 to 255(item number)
Example
Example when <NRf> is specified
:NUMERIC:HARMONICS:VALUE? 1→104.75E+OO
Example when <NRf> is omitted
:NUMERIC:HARMONICS:VALUE?→104.75E+OO,
0.9584E+00, 72.01E+00, (omit),
50.086E+00
When
"
:NUMeric:FORMat
:NUMeric:NORMAL:VALUE?→#4 (Number of
bytes, 4 digits) (Series of data bytes)
Description
• When
<NRf>
data of that item number are output.
• When
<NRf>
from 1 to the item number specified using the
"
:NUMeric:HARMonics:NUMber
output in order.
• If the item of the specified number is set to
"
NONE
" or if no numerical data exist, the item
will output error data. "
is returned when "
"
" 9.91E+37 is returned if it is set to
ASCii.
"
FLOat.
"
• In addition, if the numerical data are
erroneous (the display is "
"), "
" (infinity) is returned when
INF
"
:NUMeric:FORMat
9.9E+37 is returned if it is set to "
:NUMeric:LIST?
Function
Queries all settings related to the output of the
numerical list data during harmonic
measurement.
Syntax
:NUMeric:LIST?
Example
:NUMERIC:LIST?→:NUMERIC:LIST:
ORDER 100;SELECT ALL;ITEM U,1
4-64
A list of
"
is set to FLOat
is specified, only the numerical
is omitted, the numerical data
" command are
" (Not A Number)
NAN
" is set to
:NUMeric:FORMat
Error
" or "
--OF--
" is set to "
"
ASCii.
FLOat.
:NUMeric:LIST:ITEM
Function
Syntax
Example
:NUMeric:LIST:ORDer
Function
Syntax
Example
:NUMeric:LIST:SELect
Function
Syntax
Example
:NUMeric:LIST:VALue?
Function
Syntax
"
Example
Description
:NUMeric:NORMal?
Function
Syntax
Example
Sets the output items of the numerical list data
during harmonic measurement or queries the
current setting.
:NUMeric:LIST:ITEM {<Function>,
<Element>}
:NUMeric:LIST:ITEM?
<Function> = {U|I|P|S|Q|LAMBda|PHI|PHIU|
PHII} (See the function
selection list on page 4-32
(3))
<Element> = {<NRf>|SIGMA|SIGMB} (<NRf>=1
to 4)
:NUMERIC:LIST:ITEM U,1
:NUMERIC:LIST:ITEM?→:NUMERIC:LIST:
ITEM U,1
Sets the maximum harmonic order of the
numerical list data to output during harmonic
measurement or queries the current setting.
:NUMeric:LIST:ORDer {<NRf>|ALL}
:NUMeric:LIST:ORDer?
:NUMERIC:LIST:ORDER 100
:NUMERIC:LIST:ORDER?→:NUMERIC:LIST:
ORDER 100
Sets the output components of the numerical
list data during harmonic measurement or
queries the current setting.
:NUMeric:LIST:SELect {EVEN|ODD|ALL}
:NUMeric:LIST:SELect?
:NUMERIC:LIST:SELECT ALL
:NUMERIC:LIST:SELECT?→:NUMERIC:LIST:
SELECT ALL
Queries the numerical list data during harmonic
measurement.
:NUMeric:LIST:VALue?
:NUMERIC:LIST:VALUE?→103.58E+00,
0.00E+00,103.53E+00,0.09E+00,2.07E+00,
0.04E+00,(omit),0.01E+00,0.01E+00
The numerical data of TOTal, DC, and 1st order
to "
:NUMeric:LIST:ORDer
" are output.
Queries all settings related to the numerical
data output during normal measurement.
:NUMeric:NORMal?
:NUMERIC:NORMAL?→:NUMERIC:NORMAL:
NUMBER 8;ITEM1 URMS,1;ITEM2 UMN,1;
ITEM3 UDC,1;ITEM4 UAC,1;ITEM5 IRMS,1;
ITEM6 IMN,1;ITEM7 IDC,1;ITEM8 IAC,1