YOKOGAWA PZ4000 User Manual page 59

Power analyzer communication interface
Hide thumbs Also See for PZ4000:
Table of Contents

Advertisement

4.7 DISPlay Group
:DISPlay:WAVE:MAPPing[:MODE]
Function
Sets the waveform mapping method for the split
screen or queries the current setting.
Syntax
:DISPlay:WAVE:MAPPing[:MODE] {AUTO|
FIXed|USER}
:DISPlay:WAVE:MAPPing:MODE?
Example
:DISPLAY:WAVE:MAPPING:MODE AUTO
:DISPLAY:WAVE:MAPPING:MODE?→:DISPLAY:
WAVE:MAPPING:MODE AUTO
:DISPlay:WAVE:SVALue (Scale VALue)
Function
Turns ON/OFF the scale value display or
queries the current setting.
Syntax
:DISPlay:WAVE:SVALue {<Boolean>}
:DISPlay:WAVE:SVALue?
Example
:DISPLAY:WAVE:SVALUE OFF
:DISPLAY:WAVE:SVALUE?→:DISPLAY:WAVE:
SVALUE 0
:DISPlay:WAVE:TLABel (Trace LABel)
Function
Turns ON/OFF the waveform label display or
queries the current setting.
Syntax
:DISPlay:WAVE:TLABel {<Boolean>}
:DISPlay:WAVE:TLABel?
Example
:DISPLAY:WAVE:TLABEL ON
:DISPLAY:WAVE:TLABEL?→:DISPLAY:WAVE:
TLABEL 1
Description
The waveform labels can be set using the
"
:CHANnel<x>:LABel
:DISPlay:XY?
Function
Queries all settings related to the X-Y display.
Syntax
:DISPlay:XY?
Example
:DISPLAY:XY?→:DISPLAY:XY:XTRACE 1;
POSITION 20.000E-03,80.000E-03;
INTERPOLATE LINE
4-30
" command.
:DISPlay:XY:FFT
Function
Sets the range of the FFT waveform to be
displayed on the X-Y display or queries the
current setting.
Syntax
:DISPlay:XY:FFT {<frequency>,
<frequency>|<NRf>,<NRf>}:DISPlay:XY:FFT?
<frequency> = 0 to 2.5MHz (during the
<NRf> = 0 to 5000 (when Time Base =
External or during the harmonic
measurement mode)
Example
:DISPLAY:XY:FFT 0,200KHZ
:DISPLAY:XY:FFT?→:DISPLAY:XY:
FFT 0.000E+00,200.0E+03
Description
• Set the start point first and then the end
point.
• This command is valid when
"
:DISPlay:XY:XTRace
and the equation of "
• The range and resolution of <Frequency> is
determined from the sampling rate and the
number of FFT points.
• <NRf> is set in terms of harmonic order. The
range depends on the number of FFT points
as follows. For the procedure to set the
number of FFT points, see the
"
:MATH<x>:FFT:POINt
For 1000 points: 0 to 500
For 2000 points: 0 to 1000
For 10000 points: 0 to 5000
:DISPlay:XY:INTerpolate
Function
Sets the interpolation method of the waveform
or queries the current setting.
Syntax
:DISPlay:XY:INTerpolate {OFF|LINE}
:DISPlay:XY:INTerpolate?
Example
:DISPLAY:XY:INTERPOLATE LINE
:DISPLAY:XY:INTERPOLATE?→:DISPLAY:XY:
INTERPOLATE LINE
Description
The "
:DISPlay:WAVE:INTerpolate
can be used to make the same settings and
inquiries.
normal measurement mode,
when Time Base = Internal)
" is set to "
MATH<x>
" is set to FFT.
MATH<x>
" command.
" command
IM 253710-11E
"

Advertisement

Table of Contents
loading

Table of Contents