4.16 SETup Group

The commands in the SETup Group deal with setting the measurement mode.
These commands can be used to make the same settings and inquiries as when the SETUP key on the front
panel is pressed.
:SETup
:
RESolution
PLLSource
:SETup?
Function
Queries all settings related to the measurement
mode.
Syntax
:SETup?
Example
:SETUP?→:SETUP:MODE NORMAL;
WIRING P1W2,P1W2;RESOLUTION 5
:SETup:INITialize
Function
Initializes the settings.
Syntax
:SETup:INITialize
Example
:SETUP:INITIALIZE
Description
Resets all setup parameters except
communication settings to factory default
values.
:SETup[:MODE]
Function
Sets the measurement mode or queries the
current setting.
Syntax
:SETup[:MODE] {NORMal|HARMonics}
:SETup:MODE?
Example
:SETUP:MODE NORMAL
:SETUP:MODE?→:SETUP:MODE NORMAL
IM 253710-11E
;
MODE
<Space>
?
WIRing
<Space>
?
<Space>
?
<Space>
?
INITializ e
?
NORMal
HARMonics
P1W2
,
P1W2
P1W3
P1W3
P3W3
P3W3
P3W4
P3W4
V3A3
V3A3
NONE
<NRf>
<NRf>
EXTernal
:SETup:PLLSource
Function
Sets the PLL source during harmonic
measurement or queries the current setting.
Syntax
:SETup:PLLSource {<NRf>|EXTernal}
:SETup:PLLSource?
<NRf> = 1 to 8
EXTernal = External clock
Example
:SETUP:PLLSOURCE 1
:SETUP:PLLSOURCE?→:SETUP:PLLSOURCE 1
:SETup:RESolution
Function
Sets the number of displayed digits for
numerical data or queries the current setting.
Syntax
:SETup:RESolution {<NRf>}
:SETup:RESolution?
<NRf> = 5, 6
Example
:SETUP:RESOLUTION 5
:SETUP:RESOLUTION?→:SETUP:RESOLUTION 5
4.16 SETup Group
4
4-69