Connecting the PC to the junction box

Connecting the PC
OI/LS4000/NH3/H2O-EN
An RJ45 adapter board is provided for connecting the analyzer to a
PC.
Proceed as follows:
Step
Procedure
1
Open the junction box.
2
Connect the RJ45 adapter board to terminals 23–26 in the junction
box.
3
Connect the PC to the RJ45 adapter board using a patch cable.
Connections to the terminal strip are made as follows:
Terminal Signal
23
ETH TX+
24
ETH TX-
25
ETH RX+
26
ETH RX-
Operating instruction LS4000 for NH
Function
Ethernet TX+
Ethernet TX-
Ethernet RX+
Ethernet RX-
and H
O
3
2