ABB LS4000 Operating Instructions Manual page 104

Diode laser analyzer, version for measuring nh3 and h2o highest precision under harshest conditions
Hide thumbs Also See for LS4000:
Table of Contents

Advertisement

Performing
calibration
104
Step
Procedure
5
Click on "Next".
The Calibration menu (see page 107) is displayed.
Step
Procedure
6
Allow the calibration gas to flow through the calibration cell (flow
rate approx. 60 l/hr).
7
Observe the display of measured value. When the measured value
is stable, click on "Start".
The calibration process starts.
8
Read the measurements taken during the calibration process.
9
Observe the progress bar that displays the progress of the calibra-
tion process.
When the progress bar shows 100 %, the calibration process is
complete.
10
Click on "Done".
The Main menu is displayed.
11
Disassemble the calibration set.
Resume measuring mode.
Operating instruction LS4000 for NH
and H
O
OI/LS4000/NH3/H2O-EN
3
2