Recognizing And Resolving Errors; Error Messages In "Measuring" Mode - ABB LS4000 Operating Instructions Manual

Diode laser analyzer, version for measuring nh3 and h2o highest precision under harshest conditions
Hide thumbs Also See for LS4000:
Table of Contents

Advertisement

Recognizing and resolving errors

Error messages in "Measuring" mode

Measurement status
0x001
0x004
0x008
0x010
0x020
0x040
0x080
All others
OI/LS4000/NH3/H2O-EN
Topic
Error messages in the "Measuring" operating condition ........................... 109
Error messages in the "Malfunction" operating condition ......................... 110
Meaning and cause
LOW_TRANSMISSION_WARNING
The transmission of the laser beam is
not sufficient to evaluate the meas-
urement results.
TEMP_ERROR
The instrument has recognized an
unstable temperature in the trans-
mitter unit or receiver unit.
BEAM_BLOCK_ERROR
The laser beam does not reach the
receiver unit.
BEAM_BLOCK_WARNING
The laser beam is sometimes not able
to reach the receiver unit.
CHECKSUM_ERROR
Data transmission between the
transmitter unit and receiver unit is
disrupted.
TIMING_ERROR
The analyzer is overloaded.
SAMPLING_OVERFLOW_ERROR
Data transmission between the
transmitter unit and receiver unit is
disrupted.
Unknown or cannot be resolved on
your own
Operating instruction LS4000 for NH
How to resolve the error
 Clean all optical components
through which the laser beam
passes
 Check and correct the alignment of
the transmitter unit in relation to
the receiver unit
 If the error occurs repeatedly or
continuously: Have the error re-
solved by authorized service per-
sonnel
 Clean all optical components
through which the laser beam
passes
 Remove any foreign matter that
may be present and that may in-
terrupt the transmission of the la-
ser beam
 Check and correct the alignment of
the transmitter unit in relation to
the receiver unit
 Remove any foreign matter that
may be present and that may in-
terrupt the transmission of the la-
ser beam
 Check that the transmitter unit
and receiver unit are securely in-
stalled
 Check and repair the electrical
connection between the transmit-
ter unit and receiver unit
 Reduce the measurement period
for determining the average value
 Check and repair the electrical
connection between the transmit-
ter unit and receiver unit
 Inform authorized service person-
nel of the error code
and H
O
3
2
Page
109

Advertisement

Table of Contents
loading

Table of Contents