4.7 Connecting Sample and Reagents

4.7.1 Connecting Sample Inlet and Drain Line(s)

The sample pot fills with sample and over-flows at the top to
maintain a constant head from where sample is taken to be
measured.
B
The sample drain line(s)
routed to maintain a gravity-fed drain.
The float inside the pot contains a small magnet that operates a
reed switch. When the float is in the uppermost position the
switch is held closed. If the sample stops flowing the float drops
slowly, allowing the reed switch to open, providing a sample flow
failure indication.
Single-stream Units
Using rigid nylon tubing:
1. Connect stream 1 to sample inlet connection
(6 mm OD tubing).
2. Connect
stream
1
B
connection
(10 mm OD tubing).
10 mm
B
6 mm
A
Fig. 4.10 Connecting the Sample Inlet and Drain Line(s) –
Single Stream
20
D
F
,
and
(see Fig. 4.11) must be
A
drain
line
to
sample
Multi-stream Units
Using rigid nylon tubing:
1. Connect stream 1 to black sample inlet connection
(6 mm OD tubing).
2. Connect stream 1 drain line to black sample outlet
connection
3. Connect stream 2 to red sample inlet connection
(6 mm OD tubing).
4. Connect stream 2 drain line to red sample outlet
connection
5. Connect stream 3 to green sample inlet connection
(6 mm OD tubing).
6. Connect stream 3 drain line to green sample outlet
connection
outlet
F
10 mm
D
B
A
C
E
6 mm
Fig. 4.11 Connecting the Sample Inlet and Drain Line(s) –
Multi-stream
B
(10 mm OD tubing).
D
(10 mm OD tubing).
F
(10 mm OD tubing).
IM/AZT6CR–EN Rev. R
4 Installation