Installing the purging flanges

Fitting the purging
flanges to the
installation flanges
OI/LS4000/NH3/H2O-EN
The installation steps for the purging flanges are identical for the
transmitter unit and the receiver unit.
No.
Meaning
4x M 16 bolts
1
Installation flange
2
Purging flange O-ring
3
Purging flange
4
4x M 16 nuts
5
Install the purging flanges as follows:
Step
Procedure
1
Insert the purging flange O-ring 3 into the groove of the installa-
tion flange 2.
2
Place the purging flange 4 on the installation flange 2.
3
Screw in the flange loosely.