Selecting A Suitable Cable Gland - ABB LS4000 Operating Instructions Manual

Diode laser analyzer, version for measuring nh3 and h2o highest precision under harshest conditions
Hide thumbs Also See for LS4000:
Table of Contents

Advertisement

Selecting a suitable cable gland

Selecting a suitable
cable gland
42
Select a suitable cable gland for each cable.
The following always applies:
 Shielded cables must only be routed through metal cable glands.
 The cable diameter must fit the diameter of the cable gland.
No.
Cable
Power supply
1
P/T probes
2
Analog and digital outputs
3
Receiver unit
4
Transmitter unit
5
Operating instruction LS4000 for NH
Cable gland
M20
M20
M20
M25
M25
and H
O
OI/LS4000/NH3/H2O-EN
3
2
Cable diameter
8...15 mm
7...13 mm
7...13 mm
9...17 mm
9...17 mm