Establishing the
connections
OI/LS4000/NH3/H2O-EN
Proceed as follows:
Step
Procedure
1
Connect the supply voltage to the analyzer.
2
Connect the calibration gas to the calibration gas input.
3
Connect a discharge line to the calibration gas output.
4
Open the junction box.
5
Connect a PC to the junction box (see section Connecting a PC to
the junction box (see page 55)).
The analyzer is now ready for the calibration process.
Operating instruction LS4000 for NH
and H
O
3
2
99