ABB LS4000 Operating Instructions Manual page 73

Diode laser analyzer, version for measuring nh3 and h2o highest precision under harshest conditions
Hide thumbs Also See for LS4000:
Table of Contents

Advertisement

OI/LS4000/NH3/H2O-EN
Name
Value at 4 mA
Value at 20 mA
Digital Output 1
Digital Output 2
Source
Type
Alarm level
Prev
Cancel
Next
The layout of the outputs presented in the figure and described in
the table is an example. The actual layout is specific to each device
and can differ from this.
Operating instruction LS4000 for NH
Meaning/function
Enter the measured value that corresponds to a cur-
rent of 4 mA
Enter the measured value that corresponds to a cur-
rent of 20 mA
Digital output 1
The digital output 1 is set as an error relay.
Digital output 2
Assign the measuring channel to the digital output
Factory setting: NH3
Select the digital output operating mode (HIGH_GAS /
Error)
Factory setting: HIGH_GAS
Enter the measured value at which, if it is exceeded,
the digital output must be switched (if HIGH_GAS
operating mode is selected)
Back to previous menu
Cancel the process and exit the menu
Continue to the next menu
and H
O
3
2
73