Temporarily changing the configuration

Preface
Starting the
instrument software
Logging in
Adjusting the
configuration
Installation
procedure menu
100
Only the NH
formed using the instrument software.
Given the measuring principle, the H
not need to be calibrated. The process of checking the position of
the water line in the spectrum is explained in section Checking
absorption lines in the spectrum (see page 96).
Proceed as follows:
Step
Procedure
1
Start up the PC.
2
Start up the web browser on your computer.
3
In your web browser, type the IP address of the analyzer.
The connection to the instrument software is now established.
The Main menu is displayed.
Proceed as follows:
Step
Procedure
1
Enter the operator password.
The operator password is "xs2ls".
2
Click on "Log in".
You are now logged in as an operator.
The expanded Main menu is displayed.
For the calibration process, you must temporarily change the con-
figuration of the analyzer.
Once the calibration process is completed, the original configura-
tion must be restored.
In order to facilitate restoration of the original configuration, choose one of
the following methods:
 Make a note of the original settings before each change.
 Before each change, save a screenshot of the original settings.
Step
Procedure
1
Click on "Installation".
The Installation procedure menu (see page 63) is displayed.
2
Click on "Next" to start the configuration routine.
The Installation - Cable length menu (see page 64) is displayed.
Operating instruction LS4000 for NH
sample component needs to be calibrated; it is per-
3
and H
O
3
2
O sample component does
2
OI/LS4000/NH3/H2O-EN