Fine alignment of the purging flanges
Call up the Alignment
menu
Fine alignment of the
purging flanges
60
Proceed as follows:
Step
Procedure
1
Log on as an operator. The operator password is "xs2ls".
The Main menu is displayed.
2
Call up the Alignment menu (see page 62).
The "Alignment" menu appears.
The current values "Relative transmission" and "Absolute transmis-
sion" appear.
3
Click on "Align".
The analyzer is now ready for fine alignment.
No.
Meaning
4 stud screws
1
4 fastening nuts
2
Operating instruction LS4000 for NH
and H
O
OI/LS4000/NH3/H2O-EN
3
2