Configuration
September 2016
AMS Device Manger
1.
2.
3.
4.
5.
a.
b. If the values do not match, the current meter may be attached to the wrong loop, there
After completing the test procedure, the display returns to the loop test screen to select
another output value or to end loop testing.
6.
7.
8.
2.10

Advanced functions for HART protocol

2.10.1
Saving, recalling, and cloning configuration data
Fast Keys
Use the cloning feature of the Field Communicator or the AMS Device Manger "User
Configuration" feature to configure several Rosemount 4600 Transmitter similarly. Cloning
involves configuring a transmitter, saving the configuration data, then sending a copy of the
data to a separate transmitter. Several possible procedures exist when saving, recalling, and
cloning configuration data. For complete instructions refer to the Field Communicator
Reference Manual
Note
Do not clone an absolute pressure transmitter with sealed gauge pressure configuration data.
Do not clone a sealed gauge pressure transmitter with absolute pressure data.
18
www.aotewell.com
Right click on the device and select Diagnostics and Test then Loop test from the
menu.
Select Next after setting the control loop to manual.
Select desired analog output level. Select Next.
Select Next to acknowledge output being set to desired level.
Check the reference meter installed in the test loop to verify that it displays the
commanded output value.
If the values match, the transmitter and the loop are configured and functioning
properly.
may be a fault in the wiring, the transmitter may require an output trim, or the
reference meter may be malfunctioning.
Select End and select Next to end loop testing.
Select Next to acknowledge the loop can be returned to automatic control.
Select Finish to acknowledge the method is complete.
left arrow, 1, 2
or AMS online guides. One common method is as follows: