Emerson Rosemount 4600 Reference Manual page 25

Oil & gas panel transmitter
Hide thumbs Also See for Rosemount 4600:
Table of Contents

Advertisement

Reference Manual
00809-0100-4022, Rev GA
Field Communicator
1.
Note
If transmitter configuration has not been modified, "SAVE" option in
2.
a.
b. Select to save your configuration in either the "Internal Flash" (default) or the
c.
d. Select SAVE.
3.
4.
5.
a.
b. Select Saved Configuration
c.
6.
7.
8.
9.
Note
The transmitter receiving cloned data must have the same software version (or later) as the
original transmitter.
Configuration
www.aotewell.com
Confirm and apply configuration changes to the first transmitter.
Save the configuration data:
Select SAVE from the bottom of the Field Communicator screen.
"Configuration EM" (Configuration Expansion Module).
Enter the name for this configuration file. The default name is the transmitter tag
number.
Power the receiving transmitter and connect with Field Communicator.
Access the HART Application menu by pressing the LEFT ARROW from the
HOME/ONLINE screen.
Locate the saved transmitter configuration file.
Select Offline
Select either Internal Flash Contents or Configuration EM Content depending on
where the configuration was stored per
Use the DOWN ARROW to scroll through the list of configurations in the memory
module, and use the RIGHT ARROW to select and retrieve the desired
configuration.
Select Send to transfer the configuration to the receiving transmitter.
Select OK after the control loop is set to manual.
After the configuration has been sent, select OK to acknowledge that the loop can
be returned to automatic control. When finished, the Field Communicator informs
you of the status. Repeat
sales@aotewell.com
Step
2b.
Step 3
through
Step 9
to configure another transmitter.
Industry Automation
Configuration
September 2016
Step 2
will be disabled.
19
+86-755-8660-6182

Advertisement

Table of Contents
loading