To check tuning with the multimode source

Autotune is currently available for the G1948B Electrospray source only. To
check tuning with the multimode source, do these steps.
1 Run an Autotune with the G1948B Electrospray source installed.
2 Remove the G1948B Electrospray source and install the G1978B multimode
3 Pour the Electrospray calibrant back into its original bottle or another
4 Set the Context view to Tune in the MassHunter Workstation program.
5 Save the tune file.
Multimode Source for 6410 Triple Quad LC/MS User Guide
source.
suitable container, rinse the calibrant bottle with acetonitrile, pour the
APCI calibrant into the calibrant bottle, and attach the calibrant bottle back
onto the CDS.
a Load the autotune file that was generated with the G1948B Electrospray
source.
b Adjust the lens voltages and other tune parameters as required.
c Verify that you have sufficient abundance for the tune peaks, that the
tune peaks have peak widths of 0.7 ±0.05 amu, and that mass
assignments are within ±0.15 amu of the target mass assignments.
Basic Operation and Maintenance

To check tuning with the multimode source