6.9.4
Adjustment test with internal weight (only FEJ)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time (see chap. 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Activate function [w
(see chap. 7).
Start of the adjustment test:
The test is carried out automatically.
26
CA. 2]
Display
w CA 2
a
z0
CAL
T. KnT
CAL
T. 0
CAL
T. F. 5 .
FES/FEJ-BA-e-1132
g