Weighing/Parts Counting - KERN FES 33K0.1 IPM Operating Instruction

Fes series; fej series industrial scale
Table of Contents

Advertisement

Use the F key to switch from net weight
to gross weight or vice versa
This procedure may be repeated any
number of times (max. weighing range of
the balance).

10.2 Weighing/parts counting

With parts counting you can either count parts into a container or remove parts from
a container. To count a greater number of parts the average weight per part has to
be determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness. High reference must be selected for
small parts or parts with considerably different sizes.
The process has four steps:
• Tare the weighing container
• Determine the reference unit
• Weigh in the reference unit
• Count the items
Operation
Activate function [q SET. 2]
(see chapter 7).
The display shows the reference quantity
symbol "Pcs".
If you are using a weighing con-
tainer
FES/FEJ-BA-e-1132
Net
250z0
g
B/G
350z0
g
Display
q SET 2
Pcs
z0
45